<p>The nanoscale phase distinguishing of PCL-PB-PCL blended in epoxy resin by tapping mode atomic force microscopy</p>
</title>
<aug>
<au id="A1"><snm>Li</snm><fnm>Huiqin</fnm><insr iid="I1"/><email>hqli@sjtu.edu.cn</email></au>
<au id="A2"><snm>Sun</snm><fnm>Limin</fnm><insr iid="I1"/><email>Lmsun@sjtu.edu.cn</email></au>
<au id="A3"><snm>Shen</snm><fnm>Guangxia</fnm><insr iid="I2"/><email>gxshen@sjtu.edu.cn</email></au>
<au ca="yes" id="A4"><snm>Liang</snm><fnm>Qi</fnm><insr iid="I1"/><email>qiliang@sjtu.edu.cn</email></au>
</aug>
<insg>
<ins id="I1"><p>Instrumental Analysis Center, Shanghai Jiao Tong University, Shanghai, 200240, China</p></ins>
<ins id="I2"><p>Research Institute of Micro/Nano Science and Technology, Shanghai Jiao Tong University, Shanghai, 200240, China</p></ins>
</insg>
<source>Nanoscale Research Letters</source>
<issn>1556-276X</issn>
<pubdate>2012</pubdate>
<volume>7</volume>
<issue>1</issue>
<fpage>153</fpage>
<url>http://www.nanoscalereslett.com/content/7/1/153</url>
<xrefbib><pubidlist><pubid idtype="pmpid">22360980</pubid><pubid idtype="doi">10.1186/1556-276X-7-153</pubid></pubidlist></xrefbib>
</bibl>
<history><rec><date><day>9</day><month>10</month><year>2011</year></date></rec><acc><date><day>23</day><month>2</month><year>2012</year></date></acc><pub><date><day>23</day><month>2</month><year>2012</year></date></pub></history>
<cpyrt><year>2012</year><collab>Li et al; licensee Springer.</collab><note>This is an Open Access article distributed under the terms of the Creative Commons Attribution License (<url>http://creativecommons.org/licenses/by/2.0</url>), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.</note></cpyrt>
<kwdg>
<kwd>tapping mode AFM</kwd>
<kwd>PCL-PB-PCL</kwd>
<kwd>phase image</kwd>
<kwd>force-probe</kwd>
</kwdg>
<abs>
<sec>
<st>
<p>Abstract</p>
</st>
<p>In this work, we investigated the bulk phase distinguishing of the poly(&#949;-caprolactone)-polybutadiene-poly(&#949;-caprolactone) (PCL-PB-PCL) triblock copolymer blended in epoxy resin by tapping mode atomic force microscopy (TM-AFM). We found that at a set-point amplitude ratio (<it>r</it>
<sub>sp</sub>) less than or equal to 0.85, a clear phase contrast could be obtained using a probe with a force constant of 40 N/m. When <it>r</it>
<sub>sp </sub>was decreased to 0.1 or less, the measured size of the PB-rich domain relatively shrank; however, the height images of the PB-rich domain would take reverse (translating from the original light to dark) at <it>r</it>
<sub>sp </sub>= 0.85. Force-probe measurements were carried out on the phase-separated regions by TM-AFM. According to the phase shift angle vs. <it>r</it>
<sub>sp </sub>curve, it could be concluded that the different force exerting on the epoxy matrix or on the PB-rich domain might result in the height and phase image reversion. Furthermore, the indentation depth vs. <it>r</it>
<sub>sp </sub>plot showed that with large tapping force (lower <it>r</it>
<sub>sp</sub>), the indentation depth for the PB-rich domain was nearly identical for the epoxy resin matrix.</p>
</sec>
</abs>
</fm><bdy>
<sec>
<st>
<p>Introduction</p>
</st>
<p>Tapping mode atomic force microscopy (TM-AFM) has become a widely used technique to study the structures and properties of heterogeneous polymers at nanometer scale <abbrgrp>
<abbr bid="B1">1</abbr>
<abbr bid="B2">2</abbr>
<abbr bid="B3">3</abbr>
<abbr bid="B4">4</abbr>
<abbr bid="B5">5</abbr>
<abbr bid="B6">6</abbr>
<abbr bid="B7">7</abbr>
<abbr bid="B8">8</abbr>
<abbr bid="B9">9</abbr>
</abbrgrp>. In a TM-AFM measurement, a cantilever is forced to oscillate with the probe tip at a given amplitude (<it>A</it>
<sub>0</sub>). Then, the cantilever is brought close to the specimen and made to tap the surface with a given reduced set-point amplitude (<it>A</it>
<sub>sp</sub>). The probe-sample interaction can introduce a phase shift in the vibration with respect to that of <it>A</it>
<sub>0</sub>. So, TM-AFM measurement can obtain both height and phase images simultaneously. Height image can reflect the topographical and morphological structures, while phase images are sensitive to the physical and chemical properties of the studied material, such as stiffness, viscoelasticity, and chemical composition <abbrgrp>
<abbr bid="B3">3</abbr>
<abbr bid="B10">10</abbr>
<abbr bid="B11">11</abbr>
<abbr bid="B12">12</abbr>
</abbrgrp>.</p>
<p>However, the contrasts of the height and phase images sensitively depend on experimental conditions <abbrgrp>
<abbr bid="B1">1</abbr>
<abbr bid="B13">13</abbr>
<abbr bid="B14">14</abbr>
<abbr bid="B15">15</abbr>
<abbr bid="B16">16</abbr>
</abbrgrp> including cantilever force constant, tip shape, free amplitude <it>A</it>
<sub>0</sub>, and set-point amplitude ratio (<it>r</it>
<sub>sp</sub>) (equals to <it>A</it>
<sub>sp</sub>/<it>A</it>
<sub>0</sub>). This leads to difficulties in image interpretation for heterogeneous polymer samples. Thus, interpretation of the images has attracted considerable attention. To understand the results, a force-probe mode is performed on different materials <abbrgrp>
<abbr bid="B6">6</abbr>
<abbr bid="B13">13</abbr>
<abbr bid="B17">17</abbr>
</abbrgrp>. In this mode, the amplitude and phase shift of the tapping cantilever are measured as a function of the varied tip-sample distance in order to study the indentation response of polymer surfaces in nanoscale. The tip penetration into the compliant sample is large, while very little penetration occurred on the stiff sample <abbrgrp>
<abbr bid="B18">18</abbr>
<abbr bid="B19">19</abbr>
</abbrgrp>. If the matrix is more compliant compared with the domain regions or the domain size is much larger than the probe tip diameter (about 20 nm), the indentation difference is not affected by the tip size <abbrgrp>
<abbr bid="B1">1</abbr>
<abbr bid="B2">2</abbr>
<abbr bid="B6">6</abbr>
<abbr bid="B19">19</abbr>
</abbrgrp>. In most of the reports, the investigated polymer samples were prepared by solution casting, so enrichment which occurred at the sample surfaces would affect the results except for the scan parameters <abbrgrp>
<abbr bid="B4">4</abbr>
<abbr bid="B6">6</abbr>
<abbr bid="B20">20</abbr>
<abbr bid="B21">21</abbr>
</abbrgrp>. In this paper, we studied a new heterogeneous polymer which has a more compliant domain region with the size of about 20 nm by TM-AFM. The experiments were performed on the ultrathin section of the polymer to study the structure of the bulk material. To interpret the results, we carried out the force-probe measurement on the microdomains and the matrix.</p>
<p>In this research, the polymer blends of epoxy resin with amphiphilic poly(&#949;-caprolactone)-polybutadiene-poly(&#949;-caprolactone) (PCL-PB-PCL) triblock copolymer (chemical structure seen in Figure <figr fid="F1">1</figr>) was studied by TM-AFM as in our previous work <abbrgrp>
<abbr bid="B22">22</abbr>
</abbrgrp> which indicated that the nanostructures in the blend were formed due to the polymerization-induced microphase separation of PB subchains from the matrix, cross-linked epoxy networks, whereas the PCL remained mixed with the matrix (nanostructure seen in Figure <figr fid="F2">2</figr>). To identify the components of PCL-PB-PCL within the cross-linked epoxy resin, we carried out TM-AFM studies on the polymer blend. To help analyze the height and phase images, we performed transmission electron microscopy (TEM) measurement and force-probe measurement on the polymer blend. Our work demonstrated that to obtain true height and phase images of heterogeneous polymers and avoid artifact interference, it is important to select appropriate measurement parameters and suitable tips. Additionally, the tip indentation should be considered as well.</p>
<fig id="F1"><title><p>Figure 1</p></title><caption><p>Molecular structure of PCL-PB-PCL triblock copolymer</p></caption><text>
   <p><b>Molecular structure of PCL-PB-PCL triblock copolymer</b>.</p>
</text><graphic file="1556-276X-7-153-1" hint_layout="single"/></fig>
<fig id="F2"><title><p>Figure 2</p></title><caption><p>Schematic nanostructure of the blend of PCL-PB-PCL triblock copolymer and epoxy resin</p></caption><text>
   <p><b>Schematic nanostructure of the blend of PCL-PB-PCL triblock copolymer and epoxy resin</b>.</p>
</text><graphic file="1556-276X-7-153-2" hint_layout="double"/></fig>
</sec>
<sec>
<st>
<p>Experiments</p>
</st>
<sec>
<st>
<p>Sample preparation</p>
</st>
<p>Triblock copolymer of PCL-PB-PCL was synthesized, and the blend system of PCL-PB-PCL and epoxy resin was prepared according to the procedure reported in our previous work <abbrgrp>
<abbr bid="B22">22</abbr>
</abbrgrp>. The molecular weight of the PCL-PB-PCL was determined by hydrogen-1 nuclear magnetic resonance (<sup>1</sup>H NMR). It was found that <it>M</it>
<sub>n </sub>is 14,600 and the weight fraction of PB subchains in the PCL-PB-PCL is 35%. According to <sup>1</sup>H NMR results, the content of PCL-PB-PCL in the polymer blends is about 10% by weight. The specimen section of the polymer blends was prepared using a microtome machine (<it>ca</it>. 70 nm in thickness) and used for TEM and TM-AFM examination.</p>
</sec>
<sec>
<st>
<p>Characterization techniques</p>
</st>
<p>The TEM measurement was performed with a high-resolution TEM instrument (JEM-2010, JEOL, Tokyo, Japan) at an acceleration voltage of 120 kV. The samples were stained with OsO<sub>4 </sub>to improve the image contrast. The stained specimen section was placed in a 200 mesh copper grid for observation.</p>
<p>TM-AFM measurements were performed with an AFM instrument (Nanoscope V, Veeco, Plainview, NY, USA) at ambient condition. The commercial Si cantilevers with two types of force constant (3 and 40 N/m) were used for tapping mode measurements. The height and phase images were recorded simultaneously using the AFM instrument. Tapping mode images were taken at the fundamental resonance frequency of the Si cantilever with an ultra-sharp tip (curvature radius approximately 10 nm). The typical scanning speed was 1 Hz for all measurements.</p>
</sec>
</sec>
<sec>
<st>
<p>Results</p>
</st>
<sec>
<st>
<p>Structure of PCL-PB-PCL-blended epoxy resin film</p>
</st>
<p>To unambiguously identify the components of the blends, TEM image was taken. Figure <figr fid="F3">3</figr> shows the TEM image obtained for the ultrathin section of the PCL-PB-PCL-blended epoxy resin sample. Since the PB domains are preferentially stained with OsO<sub>4 </sub>due to the C = C double bonds, the dark regions are assigned to PB microphases, whereas the bright regions are assigned to epoxy matrix and PCL because of no stain effect for these two components <abbrgrp>
<abbr bid="B13">13</abbr>
<abbr bid="B22">22</abbr>
<abbr bid="B23">23</abbr>
</abbrgrp>. Figure <figr fid="F3">3</figr> shows that PB forms nanostructures in the system and its size is about 20 nm.</p>
<fig id="F3"><title><p>Figure 3</p></title><caption><p>The TEM image of the PCL-PB-PCL-blended epoxy resin film</p></caption><text>
   <p><b>The TEM image of the PCL-PB-PCL-blended epoxy resin film</b>.</p>
</text><graphic file="1556-276X-7-153-3" hint_layout="single"/></fig>
</sec>
<sec>
<st>
<p>Effect of cantilever stiffness on AFM image contrast</p>
</st>
<p>TM-AFM images are affected by cantilever stiffness dramatically <abbrgrp>
<abbr bid="B17">17</abbr>
<abbr bid="B24">24</abbr>
</abbrgrp>. In this research, a weaker cantilever with a force constant of 3 N/m was used. A series of phase and height images were obtained at <it>A</it>
<sub>0 </sub>= 60 nm with <it>r</it>
<sub>sp </sub>which varied from 0.1 to 0.95. Figure <figr fid="F4">4</figr> shows that no clear contrast was observed in the phase and height images obtained at <it>r</it>
<sub>sp </sub>= 0.5. Since the other images have similar contrast to Figure <figr fid="F4">4</figr>, they are not presented here.</p>
<fig id="F4"><title><p>Figure 4</p></title><caption><p>Height (a) and phase (a') images of the blended film</p></caption><text>
   <p><b>Height (a) and phase (a') images of the blended film</b>. Taken with a softer tip cantilever with a spring constant of 3 N/m at <it>A</it><sub>0 </sub>= 60 nm and <it>r</it><sub>sp </sub>= 0.5. The scanning size is 500 nm &#215; 500 nm.</p>
</text><graphic file="1556-276X-7-153-4" hint_layout="single"/></fig>
<p>Under ambient condition, the contamination layer mainly composed of water is presented on the sample surface and leads to an attractive interaction between the cantilever and the surface water layer by capillary force <abbrgrp>
<abbr bid="B1">1</abbr>
</abbrgrp>. A weaker cantilever with smaller amplitude is easy to be controlled by this attractive force. So, in order to get clear contrast for a phase image, a stiffer cantilever and a larger amplitude should be used.</p>
</sec>
<sec>
<st>
<p>Effect of <it>r</it>
<sub>sp </sub>on AFM image contrast</p>
</st>
<p>It was reported that the relative image contrast of chemically different regions depended sensitively on <it>r</it>
<sub>sp </sub>and probe tips <abbrgrp>
<abbr bid="B1">1</abbr>
<abbr bid="B14">14</abbr>
<abbr bid="B24">24</abbr>
<abbr bid="B25">25</abbr>
</abbrgrp>. According to these works, in our research, a series of height and phase images were taken at different <it>r</it>
<sub>sp </sub>(ranging from 0.95 to 0.1) using a stiffer tip with a force constant of 40 N/m.</p>
<p>Figure <figr fid="F5">5</figr> shows a series of TM-AFM height (a, b, c, d) and phase (a', b', c', d') images recorded at 30 nm of the <it>A</it>
<sub>0 </sub>and at different <it>r</it>
<sub>sp </sub>of 0.95 (a, a'), 0.85 (b, b'), 0.77 (c, c'), and 0.1 (d, d'). It was found that the height and phase image contrasts were varied with the change of <it>r</it>
<sub>sp</sub>. At weak tapping with the <it>r</it>
<sub>sp </sub>of 0.95, the image contrast differences between PB-rich and epoxy matrix are not clear both in the height (a) and phase (a') images. However, decreasing the <it>r</it>
<sub>sp </sub>value to 0.85, the image contrasts are dramatically increased (b, b', c, c'). At harder tapping with the <it>r</it>
<sub>sp </sub>of 0.1, the size of the nanostructure measured from the height (d) and phase (d') images was 25% smaller compared with that from the images taken at the <it>r</it>
<sub>sp </sub>of 0.77. An abnormal phenomenon, which was observed at <it>r</it>
<sub>sp </sub>= 0.85, was that the height (c) image contrast was completely reverse relative to any other height images, though the phase (c') image contrast looked similar to that recorded at <it>r</it>
<sub>sp </sub>= 0.77. The variations of the contrast of height and phase images at different <it>r</it>
<sub>sp </sub>were different from previous results <abbrgrp>
<abbr bid="B4">4</abbr>
<abbr bid="B7">7</abbr>
<abbr bid="B26">26</abbr>
<abbr bid="B27">27</abbr>
</abbrgrp>.</p>
<fig id="F5"><title><p>Figure 5</p></title><caption><p>Height (a-d) and phase (a'-d') images of the blended film at <b><it>A</it><sub>0 </sub></b>= 30 nm</p></caption><text>
   <p><b>Height (a-d) and phase (a'-d') images of the blended film at <b><it>A</it><sub>0 </sub></b>= 30 nm</b>. (a, a'): <it>r</it><sub>sp </sub>= 0.95; (b, b'): <it>r</it><sub>sp </sub>= 0.85; (c, c'): <it>r</it><sub>sp </sub>= 0.77; (d, d'): <it>r</it><sub>sp </sub>= 0.1. The spring constant is 40 N/m, and the scan size is 500 nm &#215; 500 nm.</p>
</text><graphic file="1556-276X-7-153-5" hint_layout="double"/></fig>
</sec>
<sec>
<st>
<p>Force-probe measurement on different domains</p>
</st>
<p>As observed in the 'Effect of <it>r</it>
<sub>sp </sub>on AFM image contrast' section, the contrast of height and phase images depended sensitively on both measurement conditions and tip-sample interaction. To investigate the factors affecting the image contrast further, force-probe measurements were performed on the nanostructure domain and the matrix region of the sample surfaces (Figures <figr fid="F6">6</figr> and <figr fid="F7">7</figr>).</p>
<fig id="F6"><title><p>Figure 6</p></title><caption><p>Amplitude (a) and phase shift (b) as a function of <b><it>z </it></b>for matrix and PB regions</p></caption><text>
   <p><b>Amplitude (a) and phase shift (b) as a function of <b><it>z </it></b>for matrix and PB regions</b>. The spring constant is 40 N/m.</p>
</text><graphic file="1556-276X-7-153-6" hint_layout="double"/></fig>
<fig id="F7"><title><p>Figure 7</p></title><caption><p>Phase shift (a) and indentation (b) as a function of <b><it>r</it><sub>sp </sub></b>for matrix and PB regions</p></caption><text>
   <p><b>Phase shift (a) and indentation (b) as a function of <b><it>r</it><sub>sp </sub></b>for matrix and PB regions</b>. The spring constant is 40 N/m.</p>
</text><graphic file="1556-276X-7-153-7" hint_layout="double"/></fig>
<p>Figure <figr fid="F6">6a</figr> shows the amplitude-<it>z </it>parameter vs. <it>A</it>(<it>z</it>) function curve obtained at <it>A</it>
<sub>0 </sub>= 30 nm for the PCL-PB-PCL and epoxy resin blend system. In Figure <figr fid="F6">6a</figr>, the solid line represents the values of the amplitude <it>A</it>
<sub>is</sub>(<it>z</it>) for an infinitely stiff sample assuming a zero indentation depth (<it>&#948;</it>). The <it>A</it>(<it>z</it>) curves observed for the PB and epoxy resin blend system lie above the <it>A</it>
<sub>is</sub>(<it>z</it>) curve because of the indentation on these samples. Since the deformation of the stiff silicon tip can be negligible compared with the sample indentation, the indentation depth <it>&#948;</it>(<it>z</it>) at a given <it>z </it>could be estimated from the equation <it>&#948;</it>(<it>z</it>) = <it>A</it>(<it>z</it>)-<it>A</it>
<sub>is</sub>(<it>z</it>) <abbrgrp>
<abbr bid="B13">13</abbr>
<abbr bid="B18">18</abbr>
</abbrgrp>. The <it>&#948;</it>(<it>r</it>
<sub>sp</sub>) curves were also obtained in a similar way. The <it>&#948;</it>(<it>r</it>
<sub>sp</sub>) curves were plotted in Figure <figr fid="F7">7b</figr>. Finally, the phase shift-<it>z </it>distance curve and the phase shift-<it>r</it>
<sub>sp </sub>curve were deduced from the amplitude-<it>z </it>and the phase angle-<it>z </it>curves, which are shown in Figures <figr fid="F6">6b</figr> and <figr fid="F6">6a</figr>, respectively.</p>
</sec>
</sec>
<sec>
<st>
<p>Discussion</p>
</st>
<p>It was found that the 'dark circle' shown in the height image and the phase image (Figure <figr fid="F5">5c, c'</figr>) was close to the nanopores shown in the TEM image. So, the morphology and the phase separation observed in TM-AFM images could be verified.</p>
<p>The phase image contrast revealed in Figure <figr fid="F5">5</figr> can be interpreted according to the phase shift-<it>r</it>
<sub>sp </sub>curve. For <it>r</it>
<sub>sp </sub>&gt; 0.85, the phase shift is negative both on PB and matrix, and the phase shift difference between PB and matrix is small, which accounts for the obscure phase contrast obtained at <it>r</it>
<sub>sp </sub>= 0.95 (shown in Figure <figr fid="F5">5a'</figr>). As <it>r</it>
<sub>sp </sub>is in the region of 0.77 to 0.85, the phase shift becomes larger on the matrix than on PB, and the largest phase shift difference between PB and matrix was observed. So, in the phase images obtained at <it>r</it>
<sub>sp </sub>= 0.77 (Figure <figr fid="F5">5c'</figr>) and at <it>r</it>
<sub>sp </sub>= 0.85 (Figure <figr fid="F5">5b'</figr>), the PB-rich region shows to be darker than the matrix region, and the phase images show maximum contrast. While the <it>r</it>
<sub>sp </sub>decreases to 0.1 (Figure <figr fid="F5">5d'</figr>), a clear phase contrast is also obtained, and the PB-rich region remains to have a dark contrast. Additionally, it was found that the size of PB-rich domains decreases compared with that shown in Figure <figr fid="F5">5b', c'</figr> in both height and phase images. In the blend, the cross-linked epoxy resin of the matrix is a thermoset polymer, and the PB polymer of the domain materials is liquid at room temperature <abbrgrp>
<abbr bid="B22">22</abbr>
</abbrgrp>. The two materials are quite different in stiffness. So, it is necessary to take the indentation depth of the probe tip into consideration. The indentation depth-<it>r</it>
<sub>sp </sub>curve (Figure <figr fid="F7">7b</figr>) reveals that the indentation depth of the PB-rich region is larger than that of the matrix when <it>r</it>
<sub>sp </sub>is more than 0.8, and it is reverse when <it>r</it>
<sub>sp </sub>is less than 0.8. For larger <it>r</it>
<sub>sp</sub>, the tip-sample interaction on the matrix of PCL mixed with epoxy will be more affected by the attractive capillary force than on the PB domains because the matrix of epoxy resin is more hydrophilic than the PB domain due to the oxygen atoms. So, the stiffer matrix shows larger indentation depth. For smaller <it>r</it>
<sub>sp</sub>, the indentation depth is nearly identical for the PB-rich domain and for the matrix, and the indentation depth is just about several nanometers. Considering that the radius size of the PB domain (approximately 10 nm) is close to the tip curvature radius (approximately 10 nm), when <it>r</it>
<sub>sp </sub>is dramatically decreased, the tip contacts both the PB domain and the matrix simultaneously. The stiff edge of the domain will prevent the tip from indenting into the compliant PB-rich region. So, the size of the domain obtained is just the size of the tip, which results in the shrinkage of the measured size of the PB domain.</p>
<p>Many studies showed that to obtain images which describe the 'true' topography of a sample surface, the images should be recorded using sufficiently high set-point amplitudes and high <it>r</it>
<sub>sp </sub>values. In this study, at <it>r</it>
<sub>sp </sub>= 0.95 to 1.0, significant tip indentation in the PB region and the matrix did not take place (see Figure <figr fid="F7">7b</figr>). We thought that at this condition, the true topography is obtained (Figure <figr fid="F5">5a</figr>), but at <it>r</it>
<sub>sp </sub>= 0.85, the height image contrast shown in Figure <figr fid="F5">5b</figr> was completely reverse. We found that in the phase shift-<it>r</it>
<sub>sp </sub>curve, at <it>r</it>
<sub>sp </sub>= 0.85, the interaction force between the tip and the sample varied from the attractive interaction on the PB into the repulsive interaction on the matrix (Figure <figr fid="F7">7a</figr>), leading to a large decrease in amplitude. Thus, the PB region looks higher than the real topography in the height image as far as the feedback mechanism is concerned. So, a height artifact is observed in Figure <figr fid="F5">5b</figr>, instead of the height image contrast inversion. While <it>r</it>
<sub>sp </sub>&#8804; 0.77, the tip-sample interaction force is repulsive for both the PB and the matrix, so the height images did not undergo reverse but influenced by the effect of tip indentation.</p>
</sec>
<sec>
<st>
<p>Conclusion</p>
</st>
<p>The above results demonstrated that it is important to select appropriate scanning parameters during TM-AFM measurement in order to get a good image with the clear morphology and topography information. The topography and nanoscale phase separation of PCL-PB-PCL-blended epoxy resin have been studied by tapping mode AFM at different <it>r</it>
<sub>sp </sub>values and with different tips. The stiffer tips are found to be desired to get clear phase shift images. According to the force-probe measurements, the optimum condition for TM-AFM measurement is 0.77 &#8804; <it>r</it>
<sub>sp </sub>&lt; 0.85. At lower <it>r</it>
<sub>sp</sub>, the size of compliant PB-rich domains decreases because the tip size is close to the dimension of PB-rich domains. At <it>r</it>
<sub>sp </sub>= 0.85, the tip-sample interaction force varies from the attractive interaction on the PB into the repulsive interaction on the matrix, which leads to the height image reverse phenomena.</p>
<p>For the reasonable interpretation of the TM-AFM phase and height images, it is necessary to take the indentation depth and the tip-sample interaction force into consideration. The present study is general and should be easily transferred for the analysis of a similar system.</p>
</sec>
<sec>
<st>
<p>Abbreviations</p>
</st>
<p>
<it>A</it>
<sub>0</sub>: given amplitude; <it>A</it>
<sub>sp</sub>: set-point amplitude; PCL-PB-PCL: poly(&#949;-caprolactone)-polybutadiene-poly(&#949;-caprolactone); <it>r</it>
<sub>sp</sub>: set-point amplitude ratio; TEM: transmission electron microscopy; TM-AFM: tapping mode atomic force microscopy.</p>
</sec>
<sec>
<st>
<p>Competing interests</p>
</st>
<p>The authors declare that they have no competing interests.</p>
</sec>
<sec>
<st>
<p>Authors' contributions</p>
</st>
<p>HL carried out the AFM measurements and drafted the manuscript. LS and GS edited the manuscript. QL participated in the analysis and guidance of the study. All authors read and approved the final manuscript.</p>
</sec>
</bdy><bm>
<ack>
<sec>
<st>
<p>Acknowledgements</p>
</st>
<p>We would like to thank Professor Sixun Zheng's research group for the polymer samples. This work was partially supported by the National Natural Science Foundation of China (No. 10974134 and 31170961) and the National Key Basic Research Program (973 Project) (2010CB933900).</p>
</sec>
</ack>
<refgrp><bibl id="B1"><title><p>Factors affecting the height and phase images in tapping mode atomic force microscopy. Study of phase-separated polymer blends of poly(ethene-co-styrene) and poly(2,6-dimethyl-1,4-phenylene oxide)</p></title><aug><au><snm>Bar</snm><fnm>G</fnm></au><au><snm>Thomann</snm><fnm>Y</fnm></au><au><snm>Brandsch</snm><fnm>R</fnm></au><au><snm>Cantow</snm><fnm>HJ</fnm></au><au><snm>Whangbo</snm><fnm>MH</fnm></au></aug><source>Langmuir</source><pubdate>1997</pubdate><volume>13</volume><fpage>3807</fpage><lpage>3812</lpage><xrefbib><pubid idtype="doi">10.1021/la970091m</pubid></xrefbib></bibl><bibl id="B2"><title><p>Characterization of the morphologies and nanostructures of blends of poly(styrene) block-poly(ethene-co-but-1-ene)-block-poly(styrene) with isotactic and atactic polypropylenes by tapping-mode atomic force microscopy</p></title><aug><au><snm>Bar</snm><fnm>G</fnm></au><au><snm>Thomann</snm><fnm>Y</fnm></au><au><snm>Whangbo</snm><fnm>MH</fnm></au></aug><source>Langmuir</source><pubdate>1998</pubdate><volume>14</volume><fpage>1219</fpage><lpage>1226</lpage><xrefbib><pubid idtype="doi">10.1021/la9711544</pubid></xrefbib></bibl><bibl id="B3"><title><p>Tapping-mode AFM studies using phase detection for resolution of nanophases in segmented polyurethanes and other block copolymers</p></title><aug><au><snm>McLean</snm><fnm>RS</fnm></au><au><snm>Sauer</snm><fnm>BB</fnm></au></aug><source>Macromolecules</source><pubdate>1997</pubdate><volume>30</volume><fpage>8314</fpage><lpage>8317</lpage><xrefbib><pubid idtype="doi">10.1021/ma970350e</pubid></xrefbib></bibl><bibl id="B4"><title><p>Understanding tapping-mode atomic force microscopy data on the surface of soft block copolymers</p></title><aug><au><snm>Wang</snm><fnm>Y</fnm></au><au><snm>Song</snm><fnm>R</fnm></au><au><snm>Li</snm><fnm>YS</fnm></au><au><snm>Shen</snm><fnm>JS</fnm></au></aug><source>Surf Sci</source><pubdate>2003</pubdate><volume>530</volume><fpage>136</fpage><lpage>148</lpage><xrefbib><pubid idtype="doi">10.1016/S0039-6028(03)00388-1</pubid></xrefbib></bibl><bibl id="B5"><title><p>Dynamic atomic force microscopy methods</p></title><aug><au><snm>Garcia</snm><fnm>R</fnm></au><au><snm>Perez</snm><fnm>R</fnm></au></aug><source>Surf Sci Rep</source><pubdate>2002</pubdate><volume>47</volume><fpage>197</fpage><lpage>301</lpage><xrefbib><pubid idtype="doi">10.1016/S0167-5729(02)00077-8</pubid></xrefbib></bibl><bibl id="B6"><title><p>Optimum topographical and morphological information in AFM tapping mode investigation of multicomponent polyethylene</p></title><aug><au><snm>Godehardt</snm><fnm>R</fnm></au><au><snm>Lebek</snm><fnm>W</fnm></au><au><snm>Adhikari</snm><fnm>R</fnm></au><au><snm>Rosenthal</snm><fnm>M</fnm></au><au><snm>Martin</snm><fnm>C</fnm></au><au><snm>Frangov</snm><fnm>S</fnm></au><au><snm>Michler</snm><fnm>G</fnm></au></aug><source>Eur Polymer J</source><pubdate>2004</pubdate><volume>40</volume><fpage>917</fpage><lpage>926</lpage><xrefbib><pubid idtype="doi">10.1016/j.eurpolymj.2004.01.034</pubid></xrefbib></bibl><bibl id="B7"><title><p>A rapid route to arrays of nanostructures in thin films</p></title><aug><au><snm>Lin</snm><fnm>ZQ</fnm></au><au><snm>Kim</snm><fnm>DH</fnm></au><au><snm>Wu</snm><fnm>XD</fnm></au><au><snm>Boosahda</snm><fnm>L</fnm></au><au><snm>Stone</snm><fnm>D</fnm></au><au><snm>LaRose</snm><fnm>L</fnm></au><au><snm>Russell</snm><fnm>TP</fnm></au></aug><source>Adv Mater</source><pubdate>2002</pubdate><volume>14</volume><fpage>1373</fpage><lpage>1376</lpage><xrefbib><pubid idtype="doi">10.1002/1521-4095(20021002)14:19&lt;1373::AID-ADMA1373&gt;3.0.CO;2-F</pubid></xrefbib></bibl><bibl id="B8"><title><p>Nanoindentation and tapping mode AFM study of phase separation in poly(ethyl acrylate-co-hydroxyethyl methacrylate) copolymer networks</p></title><aug><au><snm>Sanchez</snm><fnm>MS</fnm></au><au><snm>Mateo</snm><fnm>JM</fnm></au><au><snm>Colomer</snm><fnm>FJR</fnm></au><au><snm>Ribelles</snm><fnm>JLG</fnm></au></aug><source>Eur Polymer J</source><pubdate>2006</pubdate><volume>42</volume><fpage>1378</fpage><lpage>1383</lpage><xrefbib><pubid idtype="doi">10.1016/j.eurpolymj.2005.12.018</pubid></xrefbib></bibl><bibl id="B9"><title><p>Nanophase separation of polymers exposed to simulated bonding conditions</p></title><aug><au><snm>Ye</snm><fnm>Q</fnm></au><au><snm>Wang</snm><fnm>Y</fnm></au><au><snm>Spencer</snm><fnm>P</fnm></au></aug><source>J Biomed Mater Res Part B-Appl Biomater</source><pubdate>2009</pubdate><volume>88B</volume><fpage>339</fpage><lpage>348</lpage><xrefbib><pubid idtype="doi">10.1002/jbm.b.31047</pubid></xrefbib></bibl><bibl id="B10"><title><p>Tapping mode AFM evidence for an amorphous reticular phase in a condensation-cured hybrid elastomer: alpha, omega-dihydroxypoly(dimethylsiloxane)/poly(diethonysiloxane)/fumed silica nanoparticles</p></title><aug><au><snm>Ogoshi</snm><fnm>T</fnm></au><au><snm>Fujiwara</snm><fnm>T</fnm></au><au><snm>Bertolucci</snm><fnm>M</fnm></au><au><snm>Galli</snm><fnm>G</fnm></au><au><snm>Chiellini</snm><fnm>E</fnm></au><au><snm>Chujo</snm><fnm>Y</fnm></au><au><snm>Wynne</snm><fnm>KJ</fnm></au></aug><source>J Am Chem Soc</source><pubdate>2004</pubdate><volume>126</volume><fpage>12284</fpage><lpage>12285</lpage><xrefbib><pubidlist><pubid idtype="doi">10.1021/ja0474345</pubid><pubid idtype="pmpid" link="fulltext">15453759</pubid></pubidlist></xrefbib></bibl><bibl id="B11"><title><p>Detecting surface oxygen groups on carbon nanofibers by phase contrast imaging in tapping mode AFM</p></title><aug><au><snm>Paredes</snm><fnm>JI</fnm></au><au><snm>Martinez-Alonso</snm><fnm>A</fnm></au><au><snm>Tascon</snm><fnm>JMD</fnm></au></aug><source>Langmuir</source><pubdate>2003</pubdate><volume>19</volume><fpage>7665</fpage><lpage>7668</lpage><xrefbib><pubid idtype="doi">10.1021/la020880q</pubid></xrefbib></bibl><bibl id="B12"><title><p>AFM phase imaging of soft-hydrated samples: a versatile tool to complete the chemical-physical study of liposomes</p></title><aug><au><snm>Ruozi</snm><fnm>B</fnm></au><au><snm>Tosi</snm><fnm>G</fnm></au><au><snm>Tonelli</snm><fnm>M</fnm></au><au><snm>Bondioli</snm><fnm>L</fnm></au><au><snm>Mucci</snm><fnm>A</fnm></au><au><snm>Forni</snm><fnm>F</fnm></au><au><snm>Vandelli</snm><fnm>MA</fnm></au></aug><source>J Liposome Res</source><pubdate>2009</pubdate><volume>19</volume><fpage>59</fpage><lpage>67</lpage><xrefbib><pubidlist><pubid idtype="doi">10.1080/08982100802584071</pubid><pubid idtype="pmpid" link="fulltext">19515008</pubid></pubidlist></xrefbib></bibl><bibl id="B13"><title><p>Examination of butadiene/styrene-co-butadiene rubber blends by tapping mode atomic force microscopy. Importance of the indentation depth and reduced tip-sample energy dissipation in tapping mode atomic force microscopy study of elastomers</p></title><aug><au><snm>Bar</snm><fnm>G</fnm></au><au><snm>Ganter</snm><fnm>M</fnm></au><au><snm>Brandsch</snm><fnm>R</fnm></au><au><snm>Delineau</snm><fnm>L</fnm></au><au><snm>Whangbo</snm><fnm>MH</fnm></au></aug><source>Langmuir</source><pubdate>2000</pubdate><volume>16</volume><fpage>5702</fpage><lpage>11</lpage><xrefbib><pubid idtype="doi">10.1021/la9913699</pubid></xrefbib></bibl><bibl id="B14"><title><p>Effect of tip sharpness on the relative contributions of attractive and repulsive forces in the phase imaging of tapping mode atomic force microscopy</p></title><aug><au><snm>Bar</snm><fnm>G</fnm></au><au><snm>Brandsch</snm><fnm>R</fnm></au><au><snm>Whangbo</snm><fnm>MH</fnm></au></aug><source>Surf Sci</source><pubdate>1999</pubdate><volume>422</volume><fpage>L192</fpage><lpage>L9</lpage><xrefbib><pubid idtype="doi">10.1016/S0039-6028(98)00899-1</pubid></xrefbib></bibl><bibl id="B15"><title><p>On the factors affecting the contrast of height and phase images in tapping mode atomic force microscopy</p></title><aug><au><snm>Brandsch</snm><fnm>R</fnm></au><au><snm>Bar</snm><fnm>G</fnm></au><au><snm>Whangbo</snm><fnm>MH</fnm></au></aug><source>Langmuir</source><pubdate>1997</pubdate><volume>13</volume><fpage>6349</fpage><lpage>6353</lpage><xrefbib><pubid idtype="doi">10.1021/la970822i</pubid></xrefbib></bibl><bibl id="B16"><title><p>Phase contrast and attraction-repulsion transition in tapping mode atomic force microscopy</p></title><aug><au><snm>Chen</snm><fnm>X</fnm></au><au><snm>Roberts</snm><fnm>CJ</fnm></au><au><snm>Zhang</snm><fnm>J</fnm></au><au><snm>Davies</snm><fnm>MC</fnm></au><au><snm>Tendler</snm><fnm>SJB</fnm></au></aug><source>Surf Sci</source><pubdate>2002</pubdate><volume>519</volume><fpage>L593</fpage><lpage>L598</lpage><xrefbib><pubid idtype="doi">10.1016/S0039-6028(02)02217-3</pubid></xrefbib></bibl><bibl id="B17"><title><p>Optimizing phase imaging via dynamic force curves</p></title><aug><au><snm>Chen</snm><fnm>X</fnm></au><au><snm>Davies</snm><fnm>MC</fnm></au><au><snm>Roberts</snm><fnm>CJ</fnm></au><au><snm>Tendler</snm><fnm>SJB</fnm></au><au><snm>Williams</snm><fnm>PM</fnm></au><au><snm>Burnham</snm><fnm>NA</fnm></au></aug><source>Surf Sci</source><pubdate>2000</pubdate><volume>460</volume><fpage>292</fpage><lpage>300</lpage><xrefbib><pubid idtype="doi">10.1016/S0039-6028(00)00574-4</pubid></xrefbib></bibl><bibl id="B18"><title><p>Importance of the indentation depth in tapping-mode atomic force microscopy study of compliant materials</p></title><aug><au><snm>Bar</snm><fnm>G</fnm></au><au><snm>Delineau</snm><fnm>L</fnm></au><au><snm>Brandsch</snm><fnm>R</fnm></au><au><snm>Bruch</snm><fnm>M</fnm></au><au><snm>Whangbo</snm><fnm>MH</fnm></au></aug><source>Appl Phys Lett</source><pubdate>1999</pubdate><volume>75</volume><fpage>4198</fpage><lpage>4200</lpage><xrefbib><pubid idtype="doi">10.1063/1.125581</pubid></xrefbib></bibl><bibl id="B19"><title><p>Mapping polymer heterogeneity using atomic force microscopy phase imaging and nanoscale indentation</p></title><aug><au><snm>Raghavan</snm><fnm>D</fnm></au><au><snm>Gu</snm><fnm>X</fnm></au><au><snm>Nguyen</snm><fnm>T</fnm></au><au><snm>VanLandingham</snm><fnm>M</fnm></au><au><snm>Karim</snm><fnm>A</fnm></au></aug><source>Macromolecules</source><pubdate>2000</pubdate><volume>33</volume><fpage>2573</fpage><lpage>2583</lpage><xrefbib><pubid idtype="doi">10.1021/ma991206r</pubid></xrefbib></bibl><bibl id="B20"><title><p>Imaging of sub-surface nano particles by tapping-mode atomic force microscopy</p></title><aug><au><snm>Jiyun</snm><fnm>F</fnm></au><au><snm>Lu-Tao</snm><fnm>W</fnm></au><au><snm>Chi-Ming</snm><fnm>C</fnm></au><au><snm>Jie</snm><fnm>X</fnm></au><au><snm>Lin</snm><fnm>L</fnm></au></aug><source>Polymer</source><pubdate>2001</pubdate><volume>42</volume><fpage>2259</fpage><lpage>2262</lpage><xrefbib><pubid idtype="doi">10.1016/S0032-3861(00)00464-X</pubid></xrefbib></bibl><bibl id="B21"><title><p>Subsurface imaging of soft polymeric materials with nanoscale resolution</p></title><aug><au><snm>Spitzner</snm><fnm>EC</fnm></au><au><snm>Riesch</snm><fnm>C</fnm></au><au><snm>Magerle</snm><fnm>R</fnm></au></aug><source>Acs Nano</source><pubdate>2011</pubdate><volume>5</volume><fpage>315</fpage><lpage>320</lpage><xrefbib><pubidlist><pubid idtype="doi">10.1021/nn1027278</pubid><pubid idtype="pmpid" link="fulltext">21174404</pubid></pubidlist></xrefbib></bibl><bibl id="B22"><title><p>Nanostructured thermosetting blends of epoxy resin and amphiphilic poly(epsilon-caprolactone)-block-polybutadiene-block-poly(epsilon-caprolactone) triblock copolymer</p></title><aug><au><snm>Meng</snm><fnm>FL</fnm></au><au><snm>Zheng</snm><fnm>SX</fnm></au><au><snm>Zhang</snm><fnm>W</fnm></au><au><snm>Li</snm><fnm>HQ</fnm></au><au><snm>Liang</snm><fnm>Q</fnm></au></aug><source>Macromolecules</source><pubdate>2006</pubdate><volume>39</volume><fpage>711</fpage><lpage>719</lpage><xrefbib><pubid idtype="doi">10.1021/ma0518499</pubid></xrefbib></bibl><bibl id="B23"><title><p>Synthesis and morphology of ABC heteroarm star terpolymers of polystyrene, polybutadiene and poly(2-vinylpyridine)</p></title><aug><au><snm>Huckstadt</snm><fnm>H</fnm></au><au><snm>Gopfert</snm><fnm>A</fnm></au><au><snm>Abetz</snm><fnm>V</fnm></au></aug><source>Macromol Chem Phys</source><pubdate>2000</pubdate><volume>201</volume><fpage>296</fpage><lpage>307</lpage><xrefbib><pubid idtype="doi">10.1002/(SICI)1521-3935(20000201)201:3&lt;296::AID-MACP296&gt;3.0.CO;2-U</pubid></xrefbib></bibl><bibl id="B24"><title><p>Probing material properties of polymeric surface layers with tapping mode AFM: which cantilever spring constant, tapping amplitude and amplitude set point gives good image contrast and minimal surface damage?</p></title><aug><au><snm>Thormann</snm><fnm>E</fnm></au><au><snm>Pettersson</snm><fnm>T</fnm></au><au><snm>Kettle</snm><fnm>J</fnm></au><au><snm>Claesson</snm><fnm>PM</fnm></au></aug><source>Ultramicroscopy</source><pubdate>2010</pubdate><volume>110</volume><fpage>313</fpage><lpage>319</lpage><xrefbib><pubidlist><pubid idtype="doi">10.1016/j.ultramic.2010.01.002</pubid><pubid idtype="pmpid" link="fulltext">20133064</pubid></pubidlist></xrefbib></bibl><bibl id="B25"><title><p>Phase contrast imaging of nanocomposites and molecularly thick lubricant films in magnetic media</p></title><aug><au><snm>Bhushan</snm><fnm>B</fnm></au><au><snm>Qi</snm><fnm>J</fnm></au></aug><source>Nanotechnology</source><pubdate>2003</pubdate><volume>14</volume><fpage>886</fpage><lpage>895</lpage><xrefbib><pubid idtype="doi">10.1088/0957-4484/14/8/309</pubid></xrefbib></bibl><bibl id="B26"><title><p>Image contrast inversion of a solvent cast SEBS film</p></title><aug><au><snm>Han</snm><fnm>X</fnm></au><au><snm>Xu</snm><fnm>J</fnm></au><au><snm>Liu</snm><fnm>HL</fnm></au><au><snm>Hu</snm><fnm>Y</fnm></au></aug><source>Chin J Chem</source><pubdate>2006</pubdate><volume>24</volume><fpage>149</fpage><lpage>52</lpage><xrefbib><pubid idtype="doi">10.1002/cjoc.200690013</pubid></xrefbib></bibl><bibl id="B27"><title><p>Factors affecting phase and height contrast of diblock copolymer PS-b-PEO thin films in dynamic force mode atomic force microscopy</p></title><aug><au><snm>Wang</snm><fnm>H</fnm></au><au><snm>Djurisic</snm><fnm>AB</fnm></au><au><snm>Chan</snm><fnm>WK</fnm></au><au><snm>Xie</snm><fnm>MH</fnm></au></aug><source>Appl Surf Sci</source><pubdate>2005</pubdate><volume>252</volume><fpage>1092</fpage><lpage>1100</lpage><xrefbib><pubid idtype="doi">10.1016/j.apsusc.2005.02.034</pubid></xrefbib></bibl></refgrp>